Every Exam PrepFREE EXAM PREP

Certified Clinical Hemodialysis Technician (CCHT) Exam Study Guide

Written by Every Exam Prep Editorial TeamPublished November 16, 2026
Verified against Centers for Medicare & Medicaid Services's official sources Lead editor Vincent RuanSource and review policy
On this guide · 9 sections
Questions
150
Time limit
3h
Passing score
Standard score of 95
Exam fee
$225

From training to credential

Most people reach the CCHT through a job, not the other way round. A facility hires a technician, trains the technician, and then the clock on certification starts. This section follows that path, because the order of events shapes what you need to hand in.

The federal deadline

Under 42 CFR 494.140(e)(4), a newly hired patient care technician must be certified within 18 months of the hire date, by a state program or a national program. CMS's guidance lists the CCHT, the BONENT exam and the NNCO exam as the three national programs it approved. Until the credential is earned, the technician must have finished the facility's training program before working on patients independently, and that program has to be directed by a registered nurse and approved by the medical director and governing body.

What the training must cover

The rule lists the subjects: principles of dialysis; care of patients with kidney failure including interpersonal skills; dialysis procedures and documentation including initiation, cannulation, monitoring and termination; complications; water treatment and dialysate preparation; infection control; safety; and dialyzer reprocessing where it applies. The CCHT blueprint maps onto that list, which is why the areas of the exam read like a job description.

The application

NNCC wants a diploma or GED copy whose name matches the application, a signed verification of the training program, and the hours of hands-on training with the name of the facility where they happened. If you work as a technician, your supervisor signs for your employment. If you have been out of the field for 18 months, you must show recent classroom and hands-on retraining. The completed application goes to C-NET with the fee, and a permit follows by email.

Infection control at the chair

Infection control is where the CCHT turns from vocabulary to judgment. The exam does not ask you to recite a policy; it describes a technician at a chair and asks what comes next. The reference for nearly all of it is the CMS interpretive guidance for 42 CFR 494.30, which restates the CDC's recommendations for hemodialysis units.

Gloves and hands

Gloves are worn whenever you care for a patient or touch the patient's equipment, and they come off between patients and stations. A new, clean pair is used each time for access site care, cannulation, parenteral medication or any invasive procedure. Hand hygiene is separate from gloves: do it before an invasive procedure, immediately after gloves come off, after touching the chair-side computer and before touching the patient, and after any direct contact. Alcohol rub of 60-90% is allowed unless the hands are visibly soiled, when soap and water is required, and washing means rubbing vigorously for 15 seconds.

Medications and supplies

Prepare multiple-dose vials in a clean central area, not at the station, and deliver doses separately; common medication carts are not used. A vial labeled for single use is entered once, and leftovers from two vials are never pooled. Syringes, swabs and vials do not go in pockets. Anything unused that went to a station stays with that patient, and tape or cloth-covered cuffs that cannot be disinfected are dedicated to one patient.

Protective equipment and transducer protectors

Wear PPE during initiation and termination and when handling needles or catheters; a sleeveless apron does not count when blood might spurt. Change soiled gear at once, dialysate included. Staff do not eat, drink or smoke in the treatment area. External transducer protectors are changed between treatments and never reused, and one that gets wet is replaced immediately and checked on the machine side.

The access and the patient

The access is the patient's lifeline, and a large share of Clinical questions are about it. Think of three jobs: keep it clean, keep it visible, and notice when it changes.

Catheters and fistulas

CMS says catheters are the most common factor in bacteremia among dialysis patients, with seven times the risk of a fistula. Aseptic technique applies to every access, catheters included. Staff wear clean or sterile gloves and do hand hygiene before and after touching a catheter site or hub, check the site at every treatment, and wear a surgical mask when lumens or the exit site are open. If the site is tender or the patient has an unexplained fever, the dressing comes off so the site can be examined. Antibiotic locks are not a routine prevention measure.

Visibility

Patients must be in view of staff during treatment, and a camera does not meet that rule. If a patient wants a blanket, the access site, the line connections and the face stay uncovered, because a dislodged needle under a blanket can mean fatal blood loss within minutes. When priming fluid is cleared by letting blood fill the circuit, someone stays with the patient the whole time the venous line is open.

Noticing change

Signs of a failing fistula or graft include difficulty cannulating, trouble stopping the bleeding afterward, and changes in the pressures seen during treatment. Facilities add device-based checks such as access flow measurement, and a staff member must read the records for adverse trends. Cramping or a drop in blood pressure sends the team back to the prescription: target weight, ultrafiltration rate and dialysate sodium. A prescription is made of the treatments per week, treatment time, dialyzer, machine settings, anticoagulation and target weight.

Monthly and periodic measures

The plan of care requires albumin, body weight and hemoglobin or hematocrit at least monthly, a delivered Kt/V at least monthly aiming at 1.2 or more, and a monthly visit with a note from a physician or advanced practitioner.

Water dialysate and machines

The Technical area is the smallest, and it is the easiest to prepare for because its answers are numbers and sequences. Water is the best place to start, since a patient is exposed to many liters of it through the membrane.

Chlorine and chloramine

Treated water must have chlorine and chloramine removed, because they cause red cell hemolysis. The system needs two carbon tanks in series. The water leaving the first tank is tested; if free chlorine is above 0.5 mg/L or chloramine is above 0.1 mg/L, the second tank is tested too, and if the last tank is over the limit the facility must correct it immediately and confirm by testing, or stop treatment, and tell the medical director. Testing happens at the start of the treatment day before patients begin and again before each patient shift, or about every 4 hours if there are no shifts, on samples taken once the system has run at least 15 minutes. If the utility reports highly chlorinated water, testing moves to every 30-60 minutes. Results go on a log.

Bacteria and endotoxin

The AAMI standard the rule adopts keeps product water under 200 CFU/mL and under 2 EU/mL of endotoxin, with action levels at 50 CFU/mL and 1 EU/mL, where corrective measures must start promptly. After adverse patient reactions, cultures and endotoxin levels are drawn and the water system evaluated.

Dialysate and the machine

Before each treatment, dialysate pH is checked with an independent meter or sensitive strips, and conductivity too where the manufacturer requires it, because a concentrate mix-up can leave conductivity acceptable and pH very low. Alarms and pressure holding tests are run before each treatment. A machine that is faulty is tagged and taken out of service. A dummy drip chamber, which bypasses the air detector, is an immediate and serious threat in the treatment area. Meters are cleaned and tested daily and calibrated on the manufacturer's schedule.

Reuse

A reused dialyzer goes back only to its own patient. Hepatitis B positive patients do not receive reused dialyzers. One chemical germicide per dialyzer, bleach as cleaner aside, and a cluster of reactions tied to reuse means reuse stops until the problem is fixed. Rinse pressure is watched, since too much can rupture membranes, and blood leaks are logged.

Cleaning isolation and emergencies

Environment questions are about what happens around the patient: cleaning, isolation, waste and emergencies.

Cleaning between patients

Clean first, then disinfect: mechanical cleaning to remove debris always comes before disinfectant. Surfaces without visible blood get low-level disinfection with soap or detergent. Visible blood needs intermediate-level disinfection, a tuberculocidal disinfectant or a 1:100 bleach dilution, followed by a second application with a new cloth, and no patient should sit at the station while this is done. Scissors, clamps, stethoscopes and cuffs are cleaned between patients, and prime waste containers are emptied and disinfected after each treatment. Allow enough time between patients for the whole station, which includes the machine, the water and concentrate connections and the chair, to be disinfected.

Hepatitis B

Hepatitis B can stay infectious on a surface for at least a week. A new patient's status is known before treatment begins, or within 7 days if admitted in an emergency. A positive surface antigen means isolation in a separate room with dedicated machines and supplies, no dialyzer reuse, and staff who do not tend susceptible patients at the same time. Susceptible patients are tested monthly and vaccinated; a response is anti-HBs of at least 10 mIU/mL, tested 1-2 months after the last dose.

Emergencies

The facility's emergency plan covers fire, equipment and power failures, water supply interruption and natural disasters, and is reviewed at least every two years. Oxygen, airways, suction, a defibrillator or AED, an artificial resuscitator and emergency drugs must always be on hand, ready, and reachable in under 4 minutes. All direct care staff hold current CPR certification, and patients learn how to disconnect themselves from the machine.

Rights records and your role

Role Responsibilities is the shortest area but it is where many candidates lose easy points, because the answers are about people and paperwork rather than equipment.

Patient rights

Patients have the right to respect and dignity, to privacy and confidentiality in treatment and in their records, to information in a form they can understand, and to be told of their right to refuse or stop treatment. They must be told the facility's policies on reuse, and about grievances: they may complain personally, anonymously or through a representative, with no reprisal or loss of service. A written notice 30 days ahead precedes involuntary discharge except where others are in immediate danger. The rights must be posted with the contacts for the state agency and the ESRD network.

Records and reporting

Records are completed promptly, kept confidential except where the rule allows release, and kept for 6 years after discharge, transfer or death. A transfer sends records within 1 working day. Infection control issues go to the medical director and the quality committee, and reportable diseases are reported as the law requires. Reports of staff or patient injuries and medication errors are reviewed and trended.

The team

The assessment team is the patient, a registered nurse, the treating physician, a social worker and a dietitian. A registered nurse must be present whenever in-center patients are on treatment. The initial assessment is due within the latter of 30 days or 13 sessions, with a follow-up within 3 months; stable patients are reassessed yearly and unstable ones monthly. Hand-offs at breaks or shift change are on the blueprint as a stated activity: say what changed in pressures, access and condition before you leave a patient.

CCHT flashcards

31 cards on the highest-yield terms and rules. Grading uses spaced repetition and saves in this browser.

Card 1 of 310 mastered
Say the answer out loud before flipping.
Browse all 31 cards
  1. What does the CCHT exam cost?

    $225, which includes a $50 non-refundable application processing fee.

  2. What are the four blueprint areas and their weights?

    Clinical 48-52%, Technical 21-25%, Environment 13-17% and Role Responsibilities 10-14%.

  3. Which cognitive level carries the most weight on the CCHT?

    Application, at 63-67% of the test.

  4. What education does a CCHT applicant need?

    A high school diploma or GED, plus a training program with classroom instruction and supervised clinical experience.

  5. How long is a CCHT permit valid?

    90 days from issue. One 90-day extension is allowed, for $125, if C-NET is contacted in writing before the window closes.

  6. Is there a limit on CCHT retakes?

    No limit. A candidate who does not pass submits a new application with the full fee.

  7. How long does the CCHT credential last?

    Three years from the last day of the month in which the candidate passed.

  8. How many questions are on the CCHT exam, and how long do you have?

    150 questions in three hours. About 25 of them are unscored pilot questions (NNCC certification page and prep guide).

  9. What score passes the CCHT exam?

    A standard score of 95, which NNCC says is reached by answering about 74% of the questions correctly.

  10. What does CCHT recertification require?

    At least 3,000 hours of dialysis technician work in the three years, plus either a recertification exam or 30 contact hours of continuing education, 10 of them in nephrology.

  11. Within what time must a new patient care technician be certified?

    Within 18 months of being hired (42 CFR 494.140(e)(4)).

  12. When are new gloves required during access care?

    A new pair of clean gloves is used each time for access site care, cannulation, parenteral medication and invasive procedures.

  13. How long should hands be rubbed together when washing?

    Vigorously, for 15 seconds, with soap.

  14. What alcohol strength is acceptable for hand rub?

    60-90% alcohol, and only if the hands are not visibly soiled.

  15. May a single-use vial be punctured more than once?

    No. Sterility cannot be guaranteed after the first needle entry.

  16. When are external transducer protectors changed?

    Between each patient treatment, and immediately if they become wet. They are never reused.

  17. What does the rule say about video monitoring of patients?

    Patients must be in view of staff, and video surveillance does not meet the requirement (42 CFR 494.60(c)(4)).

  18. What is the bleach dilution for visible blood?

    1:100, which is 300-600 mg/L free chlorine, or a tuberculocidal disinfectant.

  19. What happens to a hepatitis B positive patient's dialyzer?

    It is not reused, and the patient uses dedicated machines and supplies in a separate room.

  20. What response to hepatitis B vaccine is adequate?

    An anti-HBs result of at least 10 mIU/mL, tested 1-2 months after the last primary dose.

  21. What chlorine and chloramine levels trigger testing of the second carbon tank?

    More than 0.5 mg/L free chlorine or 0.1 mg/L chloramine at the first tank.

  22. When is chlorine tested during the day?

    At the start of the treatment day before patients begin, and before each patient shift, or about every 4 hours if there are no shifts.

  23. What must be tested with an independent device before treatment?

    Dialysate pH, and conductivity where the manufacturer requires it.

  24. What are the AAMI limits for product water?

    Under 200 CFU/mL and under 2 EU/mL endotoxin, with action levels of 50 CFU/mL and 1 EU/mL.

  25. What is a dummy drip chamber and why is it dangerous?

    A fluid-filled chamber that bypasses the air detector. It risks undetected air infusion and may be used only for maintenance outside the treatment area.

  26. How often is delivered Kt/V assessed for hemodialysis patients?

    At least monthly, by calculating delivered Kt/V or an equivalent measure (42 CFR 494.80(c)(1)).

  27. What must the initial comprehensive assessment meet?

    It is due within the latter of 30 calendar days or 13 outpatient hemodialysis sessions, with a follow-up within 3 months.

  28. How long are patient records retained?

    6 years from discharge, transfer or death. A transfer needs the record sent within 1 working day.

  29. What advance notice precedes an involuntary discharge?

    Written notice 30 days ahead, except where there is an immediate threat to the health and safety of others.

  30. What emergency equipment must always be on the premises?

    Oxygen, airways, suction, a defibrillator or AED, an artificial resuscitator and emergency drugs.

  31. Who must hold current CPR certification?

    All direct patient care staff, both nurses and patient care technicians.

CCHT glossary

34 terms the CCHT exam tests, defined in plain English.

AAMI
The Association for the Advancement of Medical Instrumentation, whose dialysate and reuse standards are adopted by reference in 42 CFR Part 494.
Anti-HBs
The antibody that shows immunity to hepatitis B. A result of at least 10 mIU/mL is an adequate response to vaccination.
Arteriovenous fistula
An access made by joining an artery to a vein in the patient's own tissue. Catheters carry about seven times the bacteremia risk of a fistula, according to the CMS guidance.
Cannulation
Placing needles into a fistula or graft to start treatment. A new pair of clean gloves and hand hygiene come first.
Carbon adsorption bed
The tank that removes chlorine and chloramine from water. Two are installed in series, and the water leaving the first is the one tested.
CFU
Colony-forming unit, the measure of viable bacteria in water. Product water must have under 200 CFU/mL, with an action level of 50 CFU/mL.
Chloramine
A disinfectant in some municipal water that must be removed before the water reaches dialysis because it causes red cell hemolysis. Under 42 CFR 494.40, a chloramine reading above 0.1 mg/L after the first carbon tank means the second tank must be tested.
Conductivity
How well the dialysate conducts electricity, which reflects its dissolved ions. It is checked with an independent device because a mix-up can leave conductivity normal while pH is very low.
Dialysate
The fluid, made from treated water and concentrates, that flows past the blood in the dialyzer. Its pH and conductivity are checked with an independent device before each treatment.
Dialyzer
The filter in the extracorporeal circuit where blood and dialysate meet across a membrane. A reused dialyzer may be reused only for the same patient (42 CFR 494.50(a)(2)).
Dummy drip chamber
A fluid-filled chamber used to bypass a machine's air detector. CMS treats its presence in the treatment area as an immediate and serious threat.
Endotoxin
A bacterial product in water that can cause pyrogenic reactions in patients. The AAMI standard limits it to under 2 EU/mL, with an action level of 1 EU/mL.
ESRD network
The regional body each facility must cooperate with. Its contact details appear with the state agency's on the posted patient rights.
Exsanguination
Loss of so much blood that it can be fatal. An unnoticed needle dislodgement or line disconnection could cause it in minutes, which is why the access stays visible.
Extracorporeal circuit
The bloodlines, drip chambers and dialyzer outside the body through which the patient's blood travels during treatment.
Hand hygiene
Washing with soap and water or using an alcohol-based rub of 60-90% alcohol. Soap and water is required when hands are visibly soiled.
HBsAg
Hepatitis B surface antigen. A patient who tests positive is isolated and dialyzed on dedicated equipment.
Hemodialysis
Treatment that cleans the blood of a person with kidney failure by passing it through a dialyzer while dialysate flows past on the other side of a membrane. The CCHT exam tests the technician's part in it.
Hemolysis
Breakdown of red blood cells. Chlorine and chloramine in dialysis water can cause it.
Hemostasis
Stopping the bleeding at the needle sites after treatment. Difficulty achieving it is one of the monitoring signs of an access problem.
Heparin
The anticoagulant commonly given in the circuit to keep blood from clotting. The heparin line is clamped during dialyzer rinsing so germicide does not back up into the monitor lines.
Infiltration
Escape of a needle tip or blood into the tissue around the access. Frequent infiltrations are a cannulation technique problem that can shorten fistula survival.
Interdisciplinary team
The patient, a registered nurse, a treating physician, a social worker and a dietitian, who assess each patient and write the plan of care.
Intermediate-level disinfection
Disinfection with a tuberculocidal disinfectant or a 1:100 dilution of bleach, required where there is visible blood.
Isolation room
A separate room, with a door, for HBsAg positive patients. New facilities need one unless CMS grants a waiver.
Kt/V
A measure of dialysis dose. The plan of care aims for a hemodialysis Kt/V of at least 1.2, and delivered Kt/V is assessed at least monthly.
Patient care technician
Any person who gives direct dialysis patient care and is not another professional. Federal rules require certification within 18 months of hire.
Pilot question
A new, unscored question on the CCHT exam. About 25 of the 150 questions are pilot items.
Plan of care
The written, individualized plan the team makes to meet each patient's needs. Its implementation begins within the latter of 30 days or 13 sessions.
Standard precautions
The CDC system of infection control that treats all blood and body fluids as potentially infectious. It includes glove use, hand hygiene and personal protective equipment.
Stenosis
A narrowing in a fistula or graft that can lead to access failure. Facilities monitor accesses for it with examination, pressures and flow measurement.
Target weight
The weight the patient is meant to reach at the end of treatment. It is part of the hemodialysis prescription.
Ultrafiltration
Removal of fluid from the patient across the membrane. Cramping or falling blood pressure prompts a review of target weight, ultrafiltration rate and dialysate sodium.
Vascular access
The site where blood leaves and returns to the body: an arteriovenous fistula, an arteriovenous graft or a central venous catheter. The plan of care requires it to be monitored to prevent access failure.

Sources

  1. 1.ESRD Program Interpretive Guidance, Version 1.1 — Centers for Medicare & Medicaid Services (accessed Nov 16, 2026)
  2. 2.42 CFR Part 494 — Conditions for Coverage for End-Stage Renal Disease Facilities — Centers for Medicare & Medicaid Services (accessed Nov 16, 2026)
  3. 3.Certified Clinical Hemodialysis Technician (CCHT) — NNCC certification page — Nephrology Nursing Certification Commission (NNCC) (accessed Nov 16, 2026)
  4. 4.NNCC Certification Examination Application Booklet (revised 7/2026) — Nephrology Nursing Certification Commission (NNCC) (accessed Nov 16, 2026)
  5. 5.NHA Certifications — NHA
  6. 6.CCHT test specifications and activity statements (2021) — Nephrology Nursing Certification Commission (NNCC) (accessed Nov 16, 2026)

Official sources

The official documents our facts about this exam are taken from.

Last verified against Centers for Medicare & Medicaid Services's official sources: